Calibration Procedure
testo 350 M/XL
Calibration of Gas Sensors
! CAUTION !
Observe safety regulation/accident prevention regulations
when handling test gases.
Use test gases in well ventilated room only.
Note:
Calibration can only be performed with calibration gas of a known composition
and concentration. Gas that has exceeded its certification date should not be used.
For most accurate results, the value of the calibration gas should be between
75% to 125% of expected value. Calibrations below 50% of the expected stack
gas concentration are suspect because of multiplication of errors.
Required equipment:
Calibration gas, non-reactive regulators, non-reactive/non-absorbing tubing, a by-pass
vent for excess calibration gas, flow regulating needle valve.
Stainless Steel Flow Controlled
Regulator with Overflow Vent

Calibration Kit With Regulator

Click here for more information about portable calibration gas kits.
! CAUTION !
DO NOT PRESSURIZE THE ANALYZER GREATER THAN 12" H2O (0.5 PSI)
WITH CALIBRATION GAS AS FAULTY READINGS AND DAMAGE MAY OCCUR.
USE A BYPASS VENT.
Note:
Hydrocarbons gas should be in a carrier of O2 to facilitate calibration.
Gases like NO2 and SO2 should only be used with non-reactive,
non-absorbent materials such as stainless steel, Teflon and C-Flex.
Use single component gases to eliminate and/or identify cross interference.
Calibration time should not need to exceed 5 minutes per gas exposure
to reach stability


Calibration Procedure
Preparation
1. Ensure the analyzer is fully charged or operating on AC current
2. Ensure the ambient air is free from background gas contaminants.
Warm-up
3. Start analyzer and operate pump for 15 minutes to purge residual gases from
sensors and to fully achieve operating stability
4. Zero the analyzer by pressing the function key
"Zero" (here the second on the left).
5. Press

to enter the menu.
6. Press

to highlight "Sensors" and press
7. Highlight "Calibrate" and press
8. Select the sensor to calibrate with

and
(here CO) and press
to accept.
9. Input the indicated value of the calibration gas
(e.g. 100)
Connect to calibration gas
10. Connect the sample probe to the calibration apparatus.
11. Ensure the needle valve to the calibration gas is closed.
12. Press the "Start" function key
Note: The total calibration gas flow will be approximately 2 liters/minute. 1 liter/minute
to the analyzer; less than 1 liter/minute vented to assure adequate supply.
Adjust gas flow
13. Start the flow of the calibration gas to the analyzer by adjusting the needle valve.
14. Carefully adjust the flow to provide 1-2 liters/minute flow through the
bypass vent.
15. When the sensor reading indicates a change, time the process for 3-5 minutes
16. During the calibration this screen appears
and shows the values.
Note: Accept values At the end of the 3-5 minutes, press the "OK" function key to
adjust the sensor and to save the values into the memory.
Escape will not calibrate
(adjust) the sensor. Memory will save the values but not calibrate (adjust) the sensor.
Prepare to calibrate other sensors
17. Purge the analyzer on fresh air for 5 minutes before beginning the next gas
Printing calibration data
Note: If there is no calibration data in the sensor (e.g. sensors with a manufacturing
date before January 2003) dashes are printed instead of the target and actual values.
Only the serial number and date of adjustment are printed.
Reliable
Customer Service
Calibration, Service & Returns
Should your testo instrument require a
repair or calibration please follow
these four simple steps to insure the
best possible service and fastest
turnaround.
- Click here for PDF of "Service
Return Form" and print out form.
- Securely box the instrument in the
appropriate size and weight
cardboard box or container.
- Include with the instrument a brief
description of the problem or the
service required
- Provide your complete contact
information (name, company,
department, address, phone, fax
and e-mail) should we have any
questions or need your
authorization for extra service
or calibration.
- Ship to:
Service Department
testo, Inc.
40 White Lake Road
Sparta, NJ 07871 USA
NOTE: A Return Authorization
Number is not required!
For technical questions and troubleshooting techniques:
call 800-227-0729, ext. 129,
or ext. 132 (Monday through Friday
8 a.m. – 5 p.m.),
or e-mail
info@testo350.com
For questions on orders:
call 800-227-0729, ext. 134,
or ext. 123 (Monday through Friday
8 a.m. – 6 p.m.),
or e-mail
info@testo350.com
Contact a local
emissions distributor
If you don't know of a testo distributor
in your area, you can use our
Distributor Locator form to find one
that services your area.
to find a testo distributor near you. |
All of our customers can take
advantage of an international network
of expert service centers throughout
Europe and Asia. Each is able to
provide professional service and
calibration expertise for any Testo
instrument.
For service outside the
U.S. visit:
testo International