Range Accuracy
Resolution
Response
Time
   
O2
0 to 25% vol. < 0.2% of m.v.
0.1 vol. %
20 s (t95)
   
CO
0 to 10,000 ppm
H2 comp.
< 5 ppm (0 to 99 ppm)
< 5% of m.v. (100 to 2,000 ppm)
< 10% of m.v. (2,001 to 10,000 ppm)
1 ppm
40 s (t90)
   
CO LOW
0 to 500 ppm
H2 comp.
< 2 ppm (0 to 39.9 ppm)
< 5% of m.v. (40 to 500 ppm)
0.1 ppm
40 s (t90)
   
NO
0 to 3,000 ppm
< 5 ppm (0 to 99 ppm)
< 5% of m.v. (100 to 2,000 ppm)
< 10% of m.v. (2,001 to 3,000 ppm)
1 ppm
30 s (t90)
   
NO LOW
0 to 300 ppm
< 2 ppm (0 to 39.9 ppm)
< 5% of m.v. (300 ppm)
0.1 ppm
30 s (t90)
   
NO2
0 to 500 ppm
< 5 ppm (0 to 99 ppm)
< 5% of m.v. (500 ppm)
0.1 ppm
40 s (t90)
   
SO2
0 to 5,000 ppm
< 5 ppm (0 to 99 ppm)
< 5% of m.v. (100 to 2,000 ppm)
< 10% of m.v. (2,001 to 5,000 ppm)
1 ppm
30 s (t90)
   
H2S
0 to 300 ppm
< 2 ppm (0 to 39.9 ppm)
< 5% of m.v. (40 to 300 ppm)
0.1 ppm
35 s (t90)
   
CxHy
0.01 to 4% < 400 ppm (100 to 4,000 ppm)
< 10% of m.v. (> 4,000 ppm)
0.001 vol. %
= 10 ppm
40 s (t90)
   
CO2
0 to 50% vol. ± 0.3% vol. +1% of m.v. (0 to 25% vol.)
± 0.5% vol. +1.5% of m.v (> 25 to 50% vol.)
0.01% vol. (0 to 25% vol.)
0.01% vol. (> 25% vol.)
10 s (t90)
   
CO2
(Calculated)
0 to CO2 max vol. % Calculated from O2
0.1 vol. %
   
Differ. Press. 1
± 80" H2O < 1% m.v. (-20" to 80" H2O)
< 1% m.v. (+20" to +80" H2O)
< 0.5% m.v. (-19" to +19" H2O)
0.01" H2O
   
Differ. Press. 2
± 16" H2O < 1% m.v. (16" to 1.2" H2O)
< 1% m.v. (+16" to +1.2" H2O)
< 0.5% m.v. (-1.2" to +1.2" H2O)
0.01" H2O
   
Efficiency
0 to 100%  
0.1%
   
Flow Velocity
0 to 7900 ft/min  
10 ft/min
   
Current Voltage
0 to 20 mA ± 0.04 mA
± 0.01 mA
   
Current Voltage
0 to 10 V ± 0.01 V
± 0.01 V
   
rpm
20 to 20,000 rpm  
1 rpm
   
Temp
-40 to 2192°F ± 0.9° F (-40 to +212° F)
± 0.5% m.v. (+212 to +2,192° F)
   
             
Range
Accuracy
Resolution
Response
Time
Min. O2 Required
in Stack Gas
Response
Factor1
Methane
0.001 to 4%2
See CxHy
See CxHy
-
2% + (2x m.v. Methane)
1
Propane
0.01 to 2.1%2
See CxHy
See CxHy
-
2% + (5x m.v. Propane)
1.5
Butane
0.01 to 1.8%2
See CxHy
See CxHy
-
2% + (6.5x m.v. Butane)
2

*All information subject to change without notice.
1Lower explosion limit (LEL must be considered)
2The HC module is factory-adjusted to methane. The user can adjust the module to another gas.
 
Specifications
Dimension: 16" x 11" x 4"
Weight: 9 lbs
Storage Temp.: -4° to 122° F
Operating Temp.: 20° to 115° F
Material: ABS
Memory: 250,000 Readings
Power Required: 110 – 230, 90-260 V
or rechargeable battery
Analyzer Box Bat.: NiMH 2-3 hrs
Control Unit Bat.: 4 AA, 8 hrs
Max. Pos. Pressure: 20" H2O
Max. Neg. Pressure: 80" H2O
CO Dilution:  
Factors: 0, 2, 5, 10, 20, 40
Accuracy: Readings + 2%
Gas: Fresh Air or N2
Warranty Information:
Analyzer: 2 years
O2 cell: 1.5 years
CO, NO, NO2, SO2, CxHy, H2S: 1 year
Probe, Battery, Printer: 1 year

For probe technical data, click here.

Click here to view EPA's verification statement.