Measuring range |
Accuracy |
Resolution |
Response Time |
| O2 measurement |
0 to 25 Vol.% |
±0.2 Vol.% |
0.01 Vol.% |
< 20 seconds |
CO measurement
(H2 compensated)
|
0 to 10,000 ppm |
±10 ppm or ±10% of mv (0 to 200 ppm)
±20 ppm or ±5% of mv (201 to 2,000 ppm)
±10% of mv (2,001 to 10,000 ppm)
|
1 ppm |
< 40 seconds |
COlow measurement (H2 compensated) |
0 to 500 ppm |
±3 ppm (0 to 39.9 ppm) ±5% of mv (remaining range) |
0.1 ppm |
< 40 seconds |
| NO measurement |
0 to 3,000 ppm |
±5 ppm (0 to 99 ppm) ±5% of mv (100 to 1,999 ppm) ±10% of mv (2,000 to 3,000 ppm) |
1 ppm |
< 30 seconds |
| NOlow measurement |
0 to 300 ppm |
±3 ppm (0 to 39.9 ppm) ±5% of mv (remaining range) |
0.1 ppm |
< 30 seconds |
| NO2 measurement* |
0 to 500 ppm |
±10 ppm (0 to 199 ppm) ±5% of mv (remaining range) |
0.1 ppm |
< 40 seconds |
| SO2 measurement* |
0 to 5,000 ppm |
±10 ppm (0 to 99 ppm) ±10% of mv (remaining range) |
1 ppm |
< 40 seconds |
Temperature measurement Probe Type K (NiCr-Ni) |
-40 to 2,192°F |
32.89°F (32 to 210.2°F) ±0.5 % of mv (remaining range) |
32.18°F |
|
Draft measurement |
-0.58 to 0.58 psi |
0.0004 psi (-0.043 to 0.043 hPa) ±1.5 % of mv (remaining range) |
0.00014 psi |
|
Differential pressure measurement |
-2.9 to 2.9 psi |
0.007 psi (-0.724 to 0.724 psi) ±1.5 % of mv (remaining range) |
0.0014 psi |
|
Absolute pressure measurement |
8.7 to 16.68 psi |
0.14 psi |
0.0014 psi |
|
Calculated parameters Efficiency Flue gas loss Flue gas dewpoint |
0 to 120 % 0 to 99.9 % 211.82°F |
|
0.1 % 0.1 % 32.18°F |
|
CO2 measurement (calculation from O2) |
0 to CO2 max. |
±0.2 Vol. % |
0.1 Vol. % |
< 40 seconds |